Cáculo De Porcentajes ("Modo Porcentaje"); Parámetros - Nahita 55033210 Manual De Instrucciones

Tabla de contenido

Publicidad

CÁCULO DE PORCENTAJES ("MODO PORCENTAJE")
El modo porcentaje permite realizar medidas sobre un valor de referencia.
!
Mediante el botón "MODE" (1.4), seleccione el Modo Porcentaje (%) .
!
En pantalla se visualizará "0.00".
!
Presione el botón "CAL" (1.3).
!
Por defecto se visualizará en pantalla "--100.00--", que indica el porcentaje de
peso que corresponderá a la muestra que va a servir como referencia.
NOTA: Si desea modificar este valor, pulse los botones MODE (1.4) (para aumentarlo)
o PRINT (1.6) (para disminuirlo). Se puede seleccionar entre 50, 100, 150, 200, 250, 300, 350
y 400 %.
!
Coloque en el plato de la balanza una muestra a la que usted va a considerar con
el peso 100 % (o el que haya seleccionado).
!
Presione el botón "TARE" y tras unos segundos, visualizará en pantalla el por-
centaje correspondiente.
!
Retire la muestra del plato.
!
En pantalla se visualizará "0.00 %"
!
Coloque la sustancia que va a ser comparada sobre el plato de pesada y en pan-
talla observará la desviación resultado de la comparación del peso de la muestra original
con la muestra problema en porcentaje.
Nota: Si desea establecer una nueva referencia pulse CAL y proceda según lo explicado
anteriormente (a partir del punto 4).
PARÁMETROS
Las balanzas Nahita serie 5033 cuentan con un Software para adaptar, lo mejor posible,
las características de la balanza a las necesidades del usuario:
T
3
ABLA
F
P
(CX-Y)
UNCIÓN
ARÁMETROS
C1
C1-0
C1-1
C1-2
C2
C2-0
C2-1
C2-3
C2-3
C3
C3-0*
C3-1
Pág. 10
Manual de instrucciones 55033XXX
CASTELLANO
N
S
OMBRE
IGNIFICADO
8
16
32
1000 dígitos
2000 dígitos
5000 dígitos
10000 dígitos
ON
Cero y tara
OFF
Revisión 3 Noviembre-06
vibrations on the table in which it is situated the balance while weighing.
5. Switch the balance off by pressing On/Off key once operation has finished.
5. CALIBRATION
Whenever observing important variations in weighing or not correct weighings, balan-
ce should be calibrated. Some factors such as variation of gravity according to latitude,
change of working place as well as an inappropriate use of the equipment lead to the
necessity of recalibrate the balance.
Important!! Do not move the balance during calibration
Consult the weights you need to calibrate your balance based on its model.
T
2
Code Model
ABLE
55033210
55033220
55033230
Calibration weights should be previously calibrated.
Choose the calibration weight you need according to the model of the balance to be cali-
brated; you need a weight with the same value as maximum capacity of the balance (see
table 2). Proceed as following:
Turn the balance on by pressing On/Off key (1.2); the pan should be empty.
The balance will perform an autotest and finally stabilizes on 0.000 g.
Note: If 0.000 g does not appear on display, press TARE key (-) until it appears and
the stabilization indicator (-) lights up.
!
Press "CAL" key (1.3) to enter calibration mode.
!
CAL 0" will be displayed.
!
Press "TARE" key (1.5).
!
CAL F" will be displayed.
!
Place the calibration weight on the pan.
!
Press "TARE" key (1.5) again.
!
CAL END" will be displayed.
!
The maximum capacity of the balance will appear on display, it is also the weight of the
standard calibration weight used.
!
Removed the calibration weight from the pan.
!
0.000 g will be displayed, indicating that balance calibration has been well performed.
It is necessary to check the results of calibration:
Version 3 November-06
Instruction manual 55033XXX
Max. capacity
Calibration weight
100 g
100 g
200 g
200 g
300 g
300 g
Page 23
ENGLISH