SECTION 6—TROUBLESHOOTING
PROBLEM
Nebulizer overheats. The
nebulizer has an internal
thermal protector to pre-
vent the device from over-
heating. If the internal
temperature reaches the
preset level, the unit will
automatically shut down, and
the power LED will turn
RED.
The nebulizer does not
operate.
Medication does not nebu-
lize properly.
Nebulizer delivers no mist.
Part No 1130202
SECTION 6—TROUBLESHOOTING
Turn the unit OFF for a minimum of 40
minutes to allow the nebulizer to cool.
Check all power sources to the nebulizer.
Push the ON/OFF button again.
Check for proper connection of the tubing
adapter.
Make sure the nebulizer is in an upright position
and not tilted beyond 30º.
Make sure the nebulizer is not empty.
Clean the nebulizer as described in the manual.
Make sure the cone has been properly placed in
the nebulizer cup.
Replace the nebulizer.
Check the intake filter and replace it if it is dirty.
Make sure the cone is in the cup and properly
seated.
35