Running ß-Ketone Control Solution
Testing a ß-Ketone Quality Control Solution
1. Insert a test strip into the monitor. If monitor was off, the
screen displays all segments for 2 seconds then the blinking
blood drop symbol and "Ket" appear.
NOTE: If the strip is removed before you start the test, the screen
goes blank.
2. Press the left/right
sample is a control. (CTL is shown on the display.)
NOTE: It is important to select control solution test so the test result
does not appear to be one of your blood test results.
NOTE: If a test is not performed within 2
minutes from the insertion of the test strip, the
screen goes blank. To perform a test, take out
then replace the test strip starting from Step 1.
3. Shake the control solution vial. Discard a
trol solution onto a clean, hard, dry surface,
i.e., control cap.
4. Pick up the monitor with test strip inserted
and touch the test strip to the control solu-
tion drop.