Display
Monitor
does not
turn on after
inserting a
test strip.
Monitor
does not
begin test
countdown
after apply-
ing a blood
sample.
46
Nova Max Plus Monitor OG PN46171D1 EN.indd 46
Displays, Meanings, Actions
What it Means
upside down or not
completely in.
correctly or there is no
battery in the monitor.
sample.
monitor automatically
turned off.
damaged.
working properly.
What to Do
Insert the test strip
correctly with the Nova
name and white tip fac-
ing up and out.
Replace the battery.
Check that the battery is
-
correctly installed with
the "+" sign facing you.
Inside the USA, call 1-800-
681-7390. Outside the
USA, contact your Nova
supplier.
Repeat the test with a
new test strip.
Repeat the test with a
new test strip.
Repeat the test with a
new test strip.
After 3 attempts, call
1-800-681-7390 (inside
the USA). Outside the
USA, contact your Nova
supplier.
9/5/12 3:22 PM