Meter and Lancing Device Cleaning and Disinfecting
How to Clean and Disinfect the Meter
WARNING
Failure to follow these instructions will damage the meter and stop it from working properly.
• DO NOT clean or disinfect the meter while performing a blood glucose or control test.
• DO NOT get any moisture in slots or openings.
• DO NOT spray anything onto the meter.
• DO NOT immerse the meter in liquid.
• Always use the same product for both cleaning and disinfecting.
1
Wash hands thoroughly
with soap and water.
2
Turn the meter off and
wipe the entire meter
surface with a Super
Sani-Cloth. Carefully wipe
around the test strip slot
and other openings.
Make sure that no liquid
enters any slot or
opening.
3
A separate Super
Sani-Cloth should be
used for cleaning and
disinfection. For
disinfecting the meter, get
a new cloth and repeat
step 2, making sure that
the surface stays wet for
2 minutes.
Make sure that no
solution is seen in any
slot or opening.
4
Wash hands thoroughly
with soap and water.