Setting Up the T
The Model 8495 T
is linearly proportional to the input signal.
When the T
RANSDUCER
through the current settings for the signal input, full scale, and time constant.
The T
RANSDUCER
the default values found in the following table.
Time Constant
Full Scale
Signal Input
Zero Adjust
Span Adjust
Low Alarm
High Alarm
LEDs
Relays
Audible Alarm
®
C
ERTIFIER
The SET key is used to change (or view) the settings of the above
parameters. By pressing the SET key repeatedly, the parameters can be
changed (or viewed). Each time the up and down arrow keys () are used
to change a value the SET key must be pressed to advance to the next
parameter.
•
Full scale
The full scale can be adjusted by pressing the up and down arrow keys
(). The available values are: 0.1000 to 99990.
•
Signal input
The signal input can be adjusted by pressing the up and down arrow
keys (). The available choices are: 0-5 VDC, 0-10 VDC,
1-5 VDC, 2-10 VDC, 0-20 mA, 4-20 mA. For 0-20 mA or 4-20 mA
settings, use the provided jumper to short the MA SIG+ terminal to the
5V SIG+ terminal.
RANSDUCER
D
RANSDUCER
ISPLAY
D
& M
ISPLAY
D
& M
ISPLAY
ONITOR
Factory Settings
D
& M
ISPLAY
& M
displays an output that
ONITOR
is first turned on it will scroll
ONITOR
comes with the parameters set to
1
10000
0-5 V
0
1.000
Off
Off
Unlatched
Unlatched
Off
Off
7
ONITOR