Technical Specifications - Hach sensION+ EC 5 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

1. Technical Specifications

Subject to change without notice.
Specifications
Display
Keypad
Measuring ranges
(resolution depends on range)
Measuring error (± 1 digit)
Reproducibility (± 1 digit)
Temperature compensation
Reference temperature (RT)
Temperature coefficient (TC)
TDS conversion factor
Calibration standards
Accepted cell constant
Dimensions of the measuring instrument and ambient conditions
Operating temperature
Storage temperature
Ambient conditions
Weight
Size
Additional technical data
Energy management
Type of protection
Power supply (batteries)
Connector
Warranty
Warranty
sensION+ EC5 HACH 0710.indd ANG:3
sensION+ EC5 HACH 0710.indd ANG:3
Liquid crystal, backlit, with pictograms
Membrane, 5 keys
Conductivity
0.01 µS/cm to 500 mS/cm
Salinity
0.0 to 1999 mg/L NaCl
2.0 to 50.0 g/L NaCl
TDS
0 mg/L to 500 g/L
Temperature
–20.0 to 150.0 °C
–4.0 to 302.0 °F
Conductivity
≤ 0.5 %
Salinity
≤ 0.5 %
TDS
≤ 0.5 %
Temperature
≤ 0.2 °C (0.4 °F)
Conductivity
± 0.1 %
Salinity
± 0.1 %
TDS
± 0.1 %
Temperature
± 0.1 °C (0.1 °F)
With built-in Pt1000 temperature sensor
20 or 25 °C (68 or 77 °F). Factory settings: 25 °C (77 °F)
0.00 to 5.00%/°C (factory settings: 2%/ °C.)
0.00 to 4.44 (factory settings: 0.64)
147 µS/cm, 1413 µS/cm, 12.88 mS/cm and 111.8 mS/cm (at 25 °C/77 °F)
Between 0.05 and 50 cm
-1
(factory settings: 1.0 cm
0 °C to 50 °C (32 °F to 122 °F)
–15 °C to 65 °C (5 °F to 149 °F)
80% relative humidity (non-condensing)
300 g
186 x 73 x 38 mm
Automatic power off after 5 minutes of inactivity
IP 67
Three 1.5V AA batteries. Autonomy of over 400 hours
MP-5, five-contact multipin
3 years
-1
)
3
7/9/10 12:50:55
7/9/10 12:50:55

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Lpv3500.97.0002Lpv3560.97.0002Lpv3562.97.0002

Tabla de contenido