Two- And Three-Point Calibration - Hach sensION+ EC 5 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

6. Operation

6.3.3. Two- and three-point calibration

Follow the diagram below.
The equipment will repeat the process described for the first stan-
dard solution with the second and third.
Dip the sensor into the tube several times to homogenize the li-
quid. Do not hold it by the tube to avoid heating of the standard
solution.
Note: if the measuring instrument is not used for 5 minutes, it shuts
off automatically.
If errors occur during calibration, an error message will appear
on screen (see pg. 19).
When you switch standards, rinse the cell with deionized water.
12
sensION+ EC5 HACH 0710.indd ANG:12
sensION+ EC5 HACH 0710.indd ANG:12
50 60
50 60
50 60
0 60
6
50 60
50 60
0 60
6
50 60
50 60
50 60
0 60
6
50 60
50 60
0 60
6
EC
@25ºC
T
t
Screw the
sensor
to the 1st tube
EC
EC
μS/cm
ºC
T
T
t
t
Screw the
sensor to the
2nd tube
To exit calibration.
One-point calibration
EC
EC
μS/cm
ºC
T
T
t
t
To exit calibration
Two-point calibration
μS/cm
ºC
μS/cm
μS/cm
ºC
μS/cm
μS/cm
ºC
μS/cm
7/9/10 12:51:00
7/9/10 12:51:00

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Lpv3500.97.0002Lpv3560.97.0002Lpv3562.97.0002

Tabla de contenido