1. Introduction
Thank you for purchasing the Particle Counter instrument.
The Particle Counter displays six channels of particle sizes, down to 0.3μm, and provides fast, easy, and
accurate readings. This meter also can measure air temperature, relative humidity, and dewpoint.
It has a 2.8-inch color TFT LCD display, and an integrated 320 x 240 pixel camera, which can take
pictures(JPG) and record videos(3GP). Information can be stored internally(up to 5,000 readings) or on a
removable micro SD memory card(sold separately), and then viewed on your PC.
2. Features
• 2.8"TFT Color LCD display
• 320 x 240 pixels
• Images(JPEG) & Video(3GP)
• MicroSD(max,use 8GB)memory card
• Simultaneously measure and display 6 channel of particle sizes
• Air temperature and humidity
• Dewpoint & Wetbulb temperature
• MAX, MIN, DIF, AVG record, Date/time setup controls
• Auto Power Off
3. Specifications
Particle Counter
Channels
0.3, 0.5, 1.0, 2.5, 5.0, 10um
3
Flow Rate
0.1ft
(2.83L/min)
Counting Efficiency
50% @ 0.3 μm; 100% for particles >0.45μm
Coincidence Loss
5% at 2,000,000 particles per ft
Data Storage
5000 sample records(MicroSD card)
Count Modes
Cumulative, Differential, Concentration
HCHO Measure
Range
0.01~5.00PPM
Basic Accuracy
±5%F.S
Display resolution
0.01ppm
CO Measure
Range
10~1000PPM
Basic Accuracy
±5%F.S
Display resolution
1ppm
07/12/2017 Version No. 001