Alarm Conditions; High And Low Oxygen Concentration Alarms; Silencing An Alarm; Low Battery Alarms - Ohio Medical Corporation MiniOX 3000 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

MiniOX
3000 Oxygen Monitor
®
automatically press
to manually proceed to Monitoring Mode
NOTE: The HIGH Alarm value:
CANNOT be set equal to, or less than,
the Low Alarm value.
CAN be disabled by increasing the alarm
set point beyond 100% until "--" displays

Alarm Conditions

High and Low Oxygen Concentration
Alarms
When the MiniOX
3000 Oxygen Monitor detects an
®
oxygen concentration that exceeds the preset alarm
limit:
the red LED flashes for that alarm (High or
Low)
an audible alarm activates
the measured concentration displays

Silencing an Alarm

The MiniOX
3000 Oxygen Monitor features a silence
®
function that allows the operator to temporarily mute
the O
concentration audible alarm.
2
1. When an audible alarm activates, press the
key (labeled with a horn icon) to mute the
alarm for up to three, 30-second intervals;
press:
once for a 30-second interval
twice (within two seconds) for a
60-second interval
three times (within four seconds) for a
90-second interval
The following appears on the display:
a horn icon
a three-bar graph showing a
"countdown" of 10 seconds per bar for
each 30 second interval
NOTE:
When the audible alarm is silenced, the
appropriate visual alarm continues to
flash.
If the alarm condition is not corrected within the
preset silence interval (30, 60, or 90 seconds):
the audible alarm reactivates
If the alarm condition is corrected and then
reoccurs within this interval:
3-4
once after Step 2

Low Battery Alarms

The MiniOX
stage battery alarm that warns of depleted and
expired battery voltage:
For periodic checks on battery status (See Section 3,
Test Functions)

Sensor Indicator

During monitoring, the MiniOX
activate if:

Error Handling

The MiniOX
test:
If the microprocessor detects an error during these
routine self-tests or during monitoring:
711449 (Rev.3) 02/2016
both audible and visual alarms activate
2. To manually terminate the silence mode, press
the down arrow (labeled with a ▼ icon).
3000 Oxygen Monitor features a two-
®
The first alarm alerts the operator that the
monitor has approximately six hours of
operating time remaining:
appears on the display
an audible alarm sounds at 30-second
intervals
If the operator does not replace the battery
after this alarm, a second low battery alarm
activates when the battery is no longer able to
support monitoring.
the monitor displays
audible and visual alarms are activated
displays and "OFF"; audible and visual alarms
the oxygen sensor becomes disconnected
the cable fails or disconnects from the sensor
or instrument
the sensor membrane is perforated
the thermistor circuit opens
3000 Oxygen Monitor performs a self-
®
when it is turned ON
after a battery is installed
the display flashes "ERR" and an error code
audible and visual alarms activate the monitor
ceases to operate until the appropriate service
Section 3, Operation
and "--"
3000 Oxygen Monitor
®

Publicidad

Tabla de contenido
loading

Productos relacionados para Ohio Medical Corporation MiniOX 3000

Tabla de contenido