MiniOX
3000 Oxygen Monitor
®
Effects of Pressure, Humidity and Temperature
To ensure accurate and reliable oxygen monitoring, it is necessary to have a thorough understand-
ing of the effects of pressure, humidity, and
temperature on the sensor.
Effects of Pressure
The sensor responds to partial pressure (not per-
centage) of oxygen. Changes in barometric pressure
change the reading, even if the percent of oxygen in
the sample remains constant.
Partial pressure of oxygen (PO
oxygen (%0
) times the pressure at which the sample
2
is measured (mmHg).
PO
2
For example: at sea level, the pressure equals 760
mmHg and dry air contains 21% O
PO
= (21%) (760 mmHg)
2
PO
If you calibrate an instrument to read 21 % at 760
mmHg partial pressure and then take the instrument to
an area above sea level, a lower reading occurs due to
a lower partial pressure. For a pressure of 700 mmHg:
PO
= (21%) (700 mmHg)
2
PO
The percent reading on the instrument is derived from
the following formula:
PO
Actual
2
PO
Sea level
2
When PO
sea level is 21 %
2
(21%) (147 mmHg)
PO
Actual = (160 mmHg) = 19.3%
2
Therefore, to eliminate error due to pressure changes,
the instrument must be calibrated at the pressure in
which it is used.
Do not expose the sensor to pressure outside
B-1
CAUTION
) equals the percent of
2
= (%O
) (mmHg)
2
. Therefore:
2
= 160 mmHg
2
= 147 mmHg
2
147 mmHg
=
160 mmHg
711449 (Rev.3) 02/2016
Appendix B, Effects of Pressure, Humidity and Temperature
the range of 600 to 900 mmHg (23.62 to 35.43
inches Hg), as this may cause inaccuracies.
Effects of Humidity
The presence of humidity in an oxygen sample de-
creases the actual concentration of oxygen. Humidity
in a sample has the same effect as diluting the sample
with another gas. For example, if 100% oxygen is satu-
rated with 100% humidity, the actual concentration of
oxygen drops to 96% -97%.
As with all oxygen gas sensors, condensation on the
sensor membrane blocks the flow of oxygen, result-
ing in a lower oxygen concentration reading and an
increased response time. This is a typical problem re-
sulting from locating the sensor downstream from the
humidifier in an oxygen delivery system; clearing the
sensor face and deflector restores normal operation.
However, to avoid this problem when using the sensor
in a breathing circuit, position the sensor upstream of
the humidifier and mount the sensor with the deflector
pointing downward to prevent moisture from draining
onto the sensor membrane.
Effects of Temperature
Due to an internal thermistor (temperature variable re-
sistor), the MiniOX
3000 sensor is minimally affected
®
by temperature change. Variations in the sensor read-
ing from temperature change are less than 3% when
the instrument is calibrated and used in a monitoring
environment of 0
C to 40°C (32°F to 104°F).
o
Operating Temperature Range:
0
to 40
o
Do not use instrument outside operating
temperature range.
Do not handle the sensor unnecessarily. Body
heat can cause the sensor's thermistor to
change disproportional to the change in gas
sample temperature at the sensing electrode.
This may produce some error until thermal
equilibrium is restored.
CAUTION
C (32
to 104
F)
o
o
o
CAUTION
CAUTION