Storage
The connector should be protected from hu-
midity by the protective cap
Place the electrode in the storage protector
with Crisolyt (KCl 3M).
Life of an electrode
The average life expectancy of a pH process
electrode oscillates between 6 or 12 months,
depending on the working conditions.
Electrodes exposed to high temperatures do
not last as long. Very alkaline mediums also
reduces the life span of electrodes.
If an electrode has too short life span,
less than three months, ask us about it.
Probably there is a more suitable electrode
for the application.
Specifications
Measuring range
Operating temp.
Max. pressure
Reference element
Diaphragm
Electrolyte
Body material
Thread
Connector
Min. immersion depth
Applications
Swimming pools, drinking water, gas scrub-
bing towers, boiler waters, detoxification,
outlet waters from a waste water treatment
plant (WWTP), etc.
Limitations
Low-conductivity solutions.
Viscous or dirty solutions able to clog quickly
the diaphragm.
Warranty
CRISON guarantees this electrode against
manufacturing defects.
Validity: 6 months from shipment date.
Limitations:
• Accidental damage.
• Inadequate applications.
• Non-fulfilment of the recommendations.
• Normal wear and tear of the electrode.

CERTIFICATE OF QUALITY

The attached electrode 53 33 has passed
the quality test and follows the next speci-
fications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4...7 (at 25 ˚C) > 98%.
• Response time, pH 4...7 < 20 s.
.
0... 14 pH
0... 80 °C
6 bar
cartridge
ceramic
gel
glass
PG 13,5
S8
17 mm
GB
11