Function tests on the instrument

Checking the ready indicator
The alarm LEDs and the ready indicator in the display must flash every 60 s after
starting the instrument.
Alarm test
The alarm test must be conducted before each use.
It is part of the self-test.
(1) Briefly press the [TEST] key.
A short alarm is triggered. These include:
temporary display indicators;
alarm LEDs flash;
acoustic signal sounds briefly and
vibration alarm is triggered briefly.
Self-test
The self-test (
(1) Briefly press the [TEST] key.
In the oxygen version, the set oxygen concentration is displayed.
Calibration (
20.8%!
"
TEST
The alarm test (see above) is triggered.
(2) Press the [TEST] key again if "
Hour glass and "
(3) Feed test gas into the instrument (possible test gases
in Section 5.2).
"
OK
(4) Press the [TEST] key again.
In addition to the remaining operation time, a "
24 hours to indicate that the self-test was successful.
GB
Fig. 6) must be conducted before each use.
Section 3.6) must be carried out if the value deviates from
" "
" "
" is displayed.
GAS
?
" are displayed
GAS
" is displayed.
ALTAIR
" "
" "
TEST
GAS
TM
Operating Manual
OPERATION
" is displayed.
?
"Test gases" table
" is displayed for
19