3.6.

Calibrating the instrument

Although the instrument is maintenance-free, it must be calibrated if the self-test
has failed. Similarly, certain national regulations may stipulate the calibration of
the instrument.
For instruments used to monitor oxygen calibration must be carried out in the
following cases:
any change in the air pressure (including height above sea-level change),
any extreme changes in the atmospheric temperature,
if the self-test failed.
For instruments used to monitor toxic gases (CO and H
carried out in the following cases:
after major shocks,
any extreme changes in the atmospheric temperature,
after use under high gas concentrations,
if the self-test failed.
Calibration - Toxic gases
Make sure that the calibration is carried out in clean, non-contaminated
ambient air.
Carry out the calibration as follows (also see Fig. 7and Fig. 8):
(1) Press [TEST] key
"
TEST
(2) Press and hold [TEST] key for 3 s.
"
TEST
After 3 s, "
equalisation.
(3) During the "
equalisation.
If no fresh air equalisation is carried out, the instrument returns to
normal operating mode.
During the equalisation, the hour glass and "
If the equalisation fails ("
operating mode after 5 s.
GB
" "
" "
" is displayed.
GAS
?
" "
" is displayed.
CAL
" "
" is displayed which prompts the user to do a fresh air
FAS
?
" "
" display, press the [TEST] key to do a fresh air
FAS
?
ERR
TM
ALTAIR
" display), the instrument returns to normal
Operating Manual
OPERATION
S), calibration must be
2
" are displayed.
FAS
21