• Testing and staff training
• Testing and staff training
To check the effi ciency of the Maxi-Smart aspirator, we recommend you
carry out dynamic tests, using a pressure gauge and vacuum gauge to
verify whether the unit is serving all the surgeries satisfactorily.
Staff instruction on the use and ordinary maintenance of the suction unit
should be performed with a new, non-contaminated unit. Surgery staff
must be instructed on how to follow the display panel and the operational
phases of the Maxi-Smart unit, to understand danger warnings and carry
out maintenance operations using Puli-Jet plus new with anti-scale (A), Pulse Clea-
ner (B) and antifoaming tablets (C).
• Operation
The suction assembly (9) (through
pipe 10) creates vacuum in the
centrifugal separator (part 11 page
40).Fluid from the surgeries goes
through the manifold (2) into the
expansion tank and then through
fi lter (8) into the centrifugal sepa-
rator (11).
The centrifugal separator separates
air from liquid: the air is exhausted
outside through a pipe (4) while the
liquids go through the amalgam
separator to be drained through the
discharge pipe connected to pipe
connector (6). The centrifugal sepa-
rator (part 11 page 40) is switched
on before the turbine (part 9), this
makes it possible to discharge any
liquids that may have accumulated
in the centrifugal separator before
suction starts. When the machine
is switched off, a software setting
keeps the aspirator running for a
minimum of 13 minutes; this delay
prevents repeated stopping and
starting induced by liquid residues,
which are harmful for the system.