480 Dual Temp Operating Instructions
The 480 is a non-contact infrared thermometer, with thermocouple Probe. You can select one
Mode at a time but can change the Mode at will.
* Pressing 6 (Scan (Infrared) button), directly enters the Non-contact Infrared Thermometer
Function (IRT Mode).
* Pressing 7 (Scan (Probe) button), directly enters the Contact Thermometer Function
(COT Mode).
!
If you need to enter IRT Mode from COT Mode directly, please press 6 only after COT
is under: "Hold".
In Non-contact Infrared Thermometer function (IRT Mode)
Simply aim the 480 at the measure target with 3 and press 6 to display the surface temperature. The
distance to target ratio is 1:1 therefore the thermometer should be positioned as close to the target
as possible.
MINIMUM OR MAXIMUM MODE
he thermometer will display the minimum or maximum reading during the measurement period until the 2 button is pressed.
T
To utilize the minimum mode, please press 6 - 2 - 6 . And keep pressing 6 for measurement.
To utilize the maximum mode, please press 6 - 2 *twice - 6 . And keep pressing 6 for measurement.
LOCK MODE
The lock mode is particularly useful for continuous monitoring of temperatures. The thermometer will continuously display the temperature for up
to 60 minutes or until the 6 button is pressed.
To utilize the lock mode, please press 6 - 2 *three times- 6 .
°C OR °F Mode
To change the °C or °F mode, please press 6 - 2 *four times- 6 .
EMISSIVITY
The infrared thermometer is supplied with a default emissivity of 0.95. The emissivity can be changed for 0.10 (10E) to 1 (100E). Changes should
only be carried out by experienced personnel. For information relating to the emissivity of specific materials, please contact the nearest retailer.
Note: non-contact infrared thermometers are not recommended for use in measuring the temperature of shiny or polished metals. To change the
emissivity, please press 6 - 2 *five times- 6 for each 0.01 (1E) adjustment- 2 .
In Contact Thermometer Function (COT Mode)
Attach the 480 at the measure target with 1 and press 7 to continuously display the temperature for up to 4 minutes. After that the device will
automatically shut off to extend the battery life. Press 7 will interrupt the scanning to display the last temperature with a "Hold" wording. To
reenter scanning just press 7 again.
1. Do not twist or rotate the probe in wrong direction.
2. Over stress on probe may cause it to break.
STORAGE & CLEANING
The sensor lens is the most delicate part of the 480 and should be kept clean. Care should be taken when cleaning the lens using a soft cloth or
cotton swab with water or medical alcohol. Allow the lens to fully dry before using. Do not submerge the 480. Store at room temperature between
-4 to +149°F (-20 to +65°C).
LCD ERROR MESSAGES
The 480 incorporates visual diagnostic messages as follows:
'Hi' or 'Lo' is displayed when the temperature being measured is outside of the range of the
instrument. 'Hi' when higher than +428°F (+220°C) and 'Lo' when lower then -27°F (-33 °C) in
'Non-contact Infrared Scan' function. (In'Contact Probe Scan' function, LCD will shows the 'Hi' when
higher than +626°F (+330 °C) and 'Lo' when lower the -67°F (-55°C)
For all other error messages it is necessary to reset the 480. To reset the 480 turn the instrument off,
remove the battery and wait for a minimum of one minute, reinsert the battery and turn on. If the error
message remains, please contact the Service Department for further assistance.
BATTERIES
The 480 incorporates visual low battery indication as follows:
'Battery Low': battery needs to be replaced,
'Battery OK': measurements
are possible
3 Infrared lens
1 Probe
2 Mode
'Battery Exhausted': measurements are not
measurements are possible
When the 'Low Battery' icon indicates the battery is low, the battery should be replaced immediately with a CR2032 lithium cell. Please
!
note: It is important to turn the instrument off before replacing the battery otherwise the 480 malfunctions.
!
8 Clip
Dispose of used battery promptly.
4 Infrared
SPECIFICATION
Reading
5 Probe
Item
Reading
Measurement Range
Accuracy
6 Scan
Battery Life
(Infrared)
Dimensions
Weight
EMC/RFI
7 Scan
(Probe)
Reading may be affected if the unit is operated within a radio frequency electromagnetic field strength of
approximately 3 volts per meter, but the performance of the instrument will not be permanently affected.
Instrucciones de operación del 480 de Temperatura Doble
El 480 es un termómetro de infrarrojo sin contacto con una Sonda termocupla. Usted puede elegir un
Modo a la vez, pero puede cambiar el Modo a voluntad.
* Al oprimir 6 (botón Scan (Infrarrojo)), ingresa directamente a la Función Termómetro
Infrarrojo Sin Contacto (Modo IRT).
* Al oprimir 7 (botón Scan (Sonda)), ingresa directamente a la Función Termómetro de
Contacto (Modo COT).
!
Si usted necesita ingresar al Modo IRT desde el Modo COT directamente, por favor
oprima 6 solo una vez que COT indique: "Hold".
En la función Termómetro Infrarrojo Sin Contacto (Modo IRT)
Simplemente apunte el 480 al objetivo de medición con 3 y oprima 6 para exhibir la temperatura
de superficie. La relación entre distancia y objetivo es de 1:1, por lo tanto el termómetro debe ser
posicionado lo más cerca del objetivo posible.
MODO MÍNIMO O MÁXIMO
El termómetro exhibirá la lectura mínima o máxima durante el período de medición hasta oprimir el botón 2 .
Para utilizar el modo mínimo, por favor oprima 6 – 2 – 6 . Y siga oprimiendo 6 para la medición.
Para utilizar el modo máximo, por favor oprima 6 – 2 *dos veces – 6 . Y siga oprimiendo 6 para la medición.
MODO DE BLOQUEO
El modo de bloqueo es especialmente útil para el monitoreo continuo de temperaturas. El termómetro exhibirá continuamente la temperatura hasta
60 minutos o hasta oprimir el botón 6 .
Para utilizar el modo de bloqueo, por favor oprima 6 – 2 *tres veces – 6 .
Modo °C O °F
Para cambiar el modo °C o °F, por favor oprima 6 – 2 *cuatro veces – 6 .
EMISIVIDAD
El termómetro infrarrojo se suministra con una emisividad por defecto de 0,95. La emisividad puede ser cambiada por 0,10 (10E) a 1 (100E). Los
cambios deben ser realizados solo por personal experimentado. Para información relacionada con la emisividad de materiales específicos, por
favor contacte al distribuidor más cercano. Nota: los termómetros infrarrojos sin contacto no están recomendados para usar en la medición de la
temperatura de metales brillantes o pulidos. Para cambiar la emisividad, por favor oprima 6 – 2 *cinco veces – 6 por cada ajuste de 0,01 (1E) – 2 .
Función Termómetro en Contacto (Modo COT)
Fije el 480 al objetivo de medición con 1 y oprima 7 para exhibir continuamente la temperatura hasta 4 minutos. Luego de esto, el dispositivo se
apagará automáticamente para extender la vida de la batería. Al oprimir 7 interrumpirá el barrido para exhibir la última temperatura con una
indicación de "Hold". Para volver a ingresar al barrido, simplemente oprima 7 de nuevo.
1. No tuerza o rote la sonda en la dirección errónea.
2. Un exceso de esfuerzo sobre la sonda puede hacer que se rompa.
ALMACENAJE Y LIMPIEZA
La lente del sensor es la parte más delicada del 480 y debe ser mantenida limpia. Debe tenerse cuidado al limpiar la lente con un paño suave o
algodón con agua o alcohol medicinal. Permita que la lente se seque completamente antes de usarla. No sumerja el 480. Almacene a temperatura
ambiente entre -4 y +149°F (-20 y +65°C).
possible
Non-contact Infrared Scan function
-27 to 428°F (-33 to 220°C)
±2% of reading or ±4°F (2°C) whichever is greater
30 hours continuous use (auto power off after 15 seconds)
6.2 in x 1.68 in x .95 in (157.5mm x 42.8 mm x 24.2 mm)
2.6 oz (74 grams) including battery
Contact Probe Scan function
-67 to 626°F (-55 to 330°C)
±1% of reading or ±2°F (1°C) whichever is greater