Blue LED flashing
accompanied by
long (1s) beeping
pattern
Blue LED flashing
accompanied by
short (500ms) and
long (1s) alternate
beeping pattern
Four green LEDs
flashing together
accompanied by
long (1s) beeping
pattern
Blue LED flashing
accompanied by a
short (500ms) short
(500ms) long (1s)
beeping pattern
Blue LED flashing
accompanied by
three short (500ms)
beeping pattern
Blue LED
periodically flashing
(no pattern)
Blue LED flashing
every 0.5 seconds
Battery Not
Providing Power
(PAP Device or USB)
The
environmental
temperature is
too high for the
battery to
operate safely
Incorrect AC
adapter being
used
PAP device
connected
to battery is
incompatible
Incorrect AC
adapter being
used
PAP device
connected
to battery is
incompatible
DC output cable
is damaged
PAP device
connected
to battery is
incompatible
Maximum
battery power
output
Battery capacity
is below 5%
Battery is in
Stand-by Mode
Ensure
that
the
where there is no excess heat. Do not
rest the battery on top of radiators or
electronic devices that may emit heat.
Do not attempt to use the battery in
temperatures exceeding 40° C / 104° F
Pilot-12 Lite: use the AC adapter provided
with your Respironics DreamStation or
Respironics System One S60 or other
compatible power adapters (see section 1.3).
Pilot-24 Lite: use the AC adapter provided
with your ResMed S9 or ResMed AirSense
10 or the Medistrom AC adapter (see section
2.3).
Ensure that the PAP device you are using
is compatible with the Pilot-12 Lite or
Pilot-24 Lite (see sections 1.3 and 2.3
respectively)
Pilot-12 Lite: use the AC adapter provided
with your Respironics DreamStation or
Respironics System One S60 or other
compatible power adapters (see section 1.3)
Pilot-24 Lite: use the AC adapter provided
with your ResMed S9 or ResMed AirSense 10
or the Medistrom AC adapter (see section 2.3)
If problem persists please contact Medistrom
technical support at 1-888-407-4791 (North
America), 1-300-146-358 (Australia) or email
support@medistrom.com
Ensure that the PAP device you are using
is compatible with the Pilot-12 Lite or
Pilot-24 Lite (see sections 1.3 and 2.3
respectively)
Ensure that the cable connecting the
Pilot-12 Lite or Pilot-24 Lite to your
compatible PAP device is not damaged
Ensure that the PAP device you are using
is compatible with the Pilot-12 Lite or
Pilot-24 Lite (see sections 1.3 and 2.3
respectively)
No action required, the battery is
operating correctly. Depending on your
prescribed pressure, enabled PAP device
features and breathing rate the battery
may reach maximum power output which
is indicated by the periodically flashing
blue LED
Charge the battery as soon as possible.
The battery power level is critically low.
Not charging the battery for a prolonged
period will cause it to go into Over
Discharge State.
Press the Multi Function Button to
wake the battery from Stand-by Mode.
Disconnect
all
Plug your DC output cable into the DC
Out port on the battery and press the
Multi Function Button.For USB Devices:
Disconnect all devices and cables. Plug
your DC output cable into the DC Out
port on the battery then plug in your USB
device into the USB Out port and press the
Multi Function Button.
battery
is
devices
and
placed
cables.
15