Automated Cleaning And Disinfecting - Sendoline S1-E Manual Del Operador

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7 Conditioning
34
7.2

Automated cleaning and disinfecting ...

7.2.1
... with a DAC UNIVERSAL
We recommend using Sirona DAC UNIVERSAL for automated
Sirona DAC Universal
cleaning, disinfection, care, and unwrapped sterilization.
For further details, refer to the operating instructions supplied with the
unit.
The pushbutton chuck is manually maintained.
Procedure _ Instrument
✔ The instrument is conditioned using the DAC UNIVERSAL.
1. Check whether the instrument is clean after conditioning under good
lighting (min. 500 lux) and color rendering index (min. 80 Ra).
2. If it is still dirty, repeat the process.
 For further conditioning, the instrument should be dry and free of
residue.
Operating Instructions S1-E / S1-A
65 82 832 D3643.201.01.02.02