Chloride With Liquid Reagent - Aqualytic AL 110 Manual De Instrucciones

Ocultar thumbs Ver también para AL 110:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 140
Methods
GB
-
CL
0.0.0
Zero
!
Test
-
CL
RESULT
52

Chloride with Liquid Reagent

-
0.5 – 20 mg/l Cl
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Fill the vial with drops of the same size by holding the bottle vertically
and squeeze slowly:
20 drops KS251 (Chloride Reagent A)
Close the vial tightly with the cap and invert several times to mix the
contents.
Fill the vial with drops of the same size by holding the bottle vertically
and squeeze slowly:
20 drops KS253 (Chloride Reagent B)
Close the vial tightly with the cap and invert several times to mix the
contents.
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 5 minutes.
(Countdown can be activated, see page 43)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l Chloride.
marks
AL110_2 03/2018

Publicidad

Tabla de contenido
loading

Tabla de contenido