5. INSTALLATION

5.1 SETTING UP THE LIGHT SOURCE
Place the light source on a stable surface (cart, counter, stand, etc.).
NOTE
DO NOT block the venting grids of the light source.
Make sure the power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light source.
Make sure power cord is completely seated into the power inlet. Secure power cord by
tightening the screw with a Phillips Head screwdriver.
CAUTION
Use only cords provided with the light source.
CAUTION
Connecting and disconnecting the power cord from the light
source is the only means of connecting and disconnecting the
unit from the mains supply.
Plug the AC power cord into a wall outlet using the three (3) prong plug supplied with the
unit.
CAUTION
To prevent electric shock, connect power cords of peripheral
equipment through medical isolation transformers.
NOTE
When using medical isolation transformer, be sure to check the
transformer power ratings.
Make sure that the power cord is connected to the main power
with three prong plug (USA use UL2601-1 rated isolation
transformers and/or power strips only).
5.2 CONNECTING THE LIGHT CABLE
Connect the light cable to endoscope and then plug the light cable endfitting into the
appropriate port of the turret on the front panel.
LIT046 CUDA
(English)
Avoid places where the light source may be splashed with
liquid.
Absolutely DO NOT use in any environment with explosive or
flammable gases.
1 2
3 4
Rev. D
Page 8 of 60