5 oPeratIon

WARNING: Read the safety section and operating instructions in this manual
before use.

setuP

1. Before proceeding, ensure that all instructions in Section 4, Installation, have been
followed, and that you have read the SAFETY section of this manual.
2. Ensure that the concentrator power switch is in the OFF position. Plug the concentra-
tor's AC connector into a properly grounded power outlet.
WARNING: As a safety feature, this device has a polarized plug (one blade is wider
than the other). This plug will fit in a polarized outlet only one way. If the plug
does not fit completely in the outlet, reverse the plug. DO NOT attempt to defeat
this safety feature. Contact a qualified electrician or your Graham-Field authorized
distributor if the plug does not fit in the polarized outlet.
turnIng the ConCentrator on
Note: For use with a humidifier bottle, please
see OperatiOn with a humidifier bOt-
tle (if sO prescribed) on the following
page.
1. Connect the oxygen tubing securely to
the oxygen outlet; connect the other end
to the patient's cannula or mask outlet
(these may already be connected).
2. Move the concentrator's ON/OFF switch
to the ON position to turn on concentra-
tor power.
Model JB0180: The green (power)
LED will illuminate.
Model JB0180S: the yellow low oxygen LED will illuminate for approximately
three minutes while the system performs an Oxygen Purity test, then the green
(power) LED and appropriate O
Ensure that oxygen is flowing through the tubing before proceeding.
WARNING: If the flow rate on the flowmeter ever falls below 0.5 LPM, check tubing
or accessories for blocked or kinked tubing or for a defective humidifier bottle.
3. Turn the flowmeter knob to center the flowmeter ball on the LPM prescribed. The
flowmeter can be adjusted only after the unit is powered on.
WARNING: Do not change LPM setting unless prescribed by a physician.
Note: Oxygenation times and flow rate ranges are prescribed by your physician.
JB0180-INS-LAB-RevA08
Optium
Oxygen Concentrator
connections
LED will illuminate.
2
oxygen
outlet
tubing
nasal
mask
cannula
10