9 sPeCIfICatIons
ItEM
Voltage
Power consumption
Flow range
Oxygen concentration *
Oxygen outlet pressure (no flow)
Operating pressure
Environment Temperature
Humidity
Atmospheric
pressure
Altitude
Sound level
Safety system
Dimensions (W x D x H)
Weight
Pressure relief valve (safety feature)
LED indicators
* Oxygen concentrator levels are for operation within specific temperature and altitude ranges.
Operation outside specified ranges may decrease oxygen concentration levels.
Outlet pressure may need adjustment to compensate for higher altitude.
Note: Specifications are subject to change without notice.
JB0180-INS-LAB-RevA08
SPECIFICAtIONS
115VAC - 60 Hz
300 W (average)
0.5 to 5 LPM
93% ±3% @0.5 to 5 LPM
8.5 psi ±0.5 psi (58.6 kPa ±3.4 kPa)
14 to 26 psi (96.5 kPa to 179.3 kPa) @5 LPM
Operating
50°F to 95°F (10°C to 35°C)
Storage/Transport -40°F to 158°F (-40°C to 70°C)
Operating/
10~100% RH non-condensing
Storage/Transport
Storage/Transport 7.3 to 15.4 psi (50 to 106 kPa)
≤7,500 ft. (≤2,286m)
60Hz
a) Current overload
b) High temperature compressor shutdown
c) High pressure alarm with compressor shutdown
d) Low pressure alarm with compressor shutdown
e) Power loss alarm
f) Model JB0180S only: OPI (Oxygen Percentage Indicator)
oxygen system alarm
14.2 x 13 x 24.8 inches (36 x 33 x 63 cm)
33 lb (15kg)
35 psi ±3 psi (241.3 kPa ±20.7 kPa)
Please see previous LED indicator table
≤45 dBa
0