13. OPERATING ANOMALIES AND TROUBLESHOOTING
Display
No initial reset
The sensor does not
switch on
No display evolution or
display inconsistent
Linearity or precision
problem
14.
REMOTE DISPLAY OPTION
The dynafor™ LLX1 dynamometers can communicate with a type LLX1 remote display.
When an LLX1 dynamometer is radio-linked with an LLX1 display, the units symbol flashes once per
second. For use of the display, refer to the manual supplied with the display.
"IMPORTANT"
Possible causes
Tare Function enabled.
Permanent deformation of the
sensor following a handling
error; excessive overload or
compression.
Dead batteries.
Electronic fault.
Sensor or sensor electronics
malfunction.
Sensor or sensor electronics
malfunction.
Sensor subjected to
compression or torsion load.
Negative unbalance of gauge
bridge.
For information, the radio range measured in laboratory is 40 m when the
front side of the sensor is pointed toward the back side of the display.
Caution ! This range can change in other situations, due in particular to:
- the presence of obstacles.
- the existence of electromagnetic interference.
- certain atmospheric conditions.
If any trouble is observed, or for special applications, contact the Tractel
network.
13
Solutions
Disable the Tare function and
display the "GROS" stress
value.
The appliance should be
checked by the manufacturer
before you continue using.
Change batteries.
Contact the after-sales service.
Reinitialise : Switch off, then
switch on the sensor.
In the event of persistent
malfunction, contact the after
sales service.
Contact the after-sales service.
Eliminate the compression load
on the sensor.
Contact the after-sales service.