Hameg Instruments HM8018 Manual Del Usuario página 11

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

L-C Meter HM8018
I 24 Measurement Ranges
I Max. Resolution: 0.1pF - 0.1µH - 0.01Ω Ω Ω Ω Ω - 0.01µS
I 3 Test Frequencies (160kHz, 1.6kHz, 16kHz)
I 4-Point Measurement Technology
I Basic Accuracy 0.5%
I Internal Bias for Electrolytic Capacitors
I Measures both Series and Parallel
Components
The LC Meter HM8018 offers a number of
features uncommon for an instrument of its
price range. Measured values are displayed on
a 3½ - digit LED with a basic accuracy of
0.5%. Measurements are performed at a 2
per second rate. Due to use of 3 oscillator
frequencies, which automatically switch upon
range selection, the instrument always opera-
tes at the optimum test frequency.
Besides L and C measurements, series and
parallel components of inductances and
capacitances can be determined with the
HM8018, providing easy evaluation of the
quality or dissipation factors of the com-
Specifications
(Reference Temperature: 23°C ±1°C)
Operating Modes:
C-Measurement
L-Measurement
Series Inductance L, Shunt Capacitance C
Series Resistance R, Shunt Conductance G
Ranges:
L : 200µ to 200H
Rs: 20Ω
to 200kΩ
C : 200pF to 200µF
G : 20µS to 200mS
Max. Resolution: 0.1pF
0.1µH
0.01Ω
0.01µS
Measuring Frequencies: (Sinewave signal)
approx. 160Hz, 1.6kHz, 16kHz
−1
(ω = 10
3
, 10
4
,10
5
s
)
Measuring Voltage: max. 1V
Measuring Current: max. 36mA (rms)
Power dissipation during measurement:
max. 3.2mW
Accuracy: ±0.5% o.v.
+(3 Digit+0.5pF/0.5µH/10mΩ/0.01µS)]
Error ≤1% for separation of real component
and imaginary part up to 45° phase angle
Values without tolerances are meant to be guidelines and represent characteristics of the average
instrument.
12
(7 decades)
(5 decades)
(7 decades)
(5 decades)
pp
1)
ponents under test. A special measurement
technique allows separation of real and imaginary
components up to a phase angle of 45° with
an accuracy of
resolution, 10mΩ Ω Ω Ω Ω / 0.01µS ensures precise and
reliable test results.
The accuracy and operability of the HM8018
make it ideal for use in virtually every area of
electronic technology, including quality control
and laboratory applications. The low number
of operating controls ensures that even non-
technical personnel can perform accurate
measurements after only brief instruction.
Display:
3½-digit 7 segment LED-display
Measuring Rate: 2 measurements/second
Principle of Measurement: 2- or 4-Point
Measurement
General Information:
Inputs short-circuit-proof and overvoltage
protected
up to 100V for a short time at an energy
consumption of
charged to 100V).
Polarization voltage for C-measurement: 2V
Zero calibration for display
Compensation of probe capacitance (HZ18)
An AC voltage proportional to display
is available at the rear BNC connector
(HM8001)
Supply voltages (from HM8001):
+5V/200mA
−13V/130mA
+13V/130mA
(∑ = 4.5W)
Operating conditions: 0°C to +40°C
max. relative humidity: 80%
Dimensions (without 22-pin flat connector):
W135, H68, D228mm
Weight: approx. 650g
1)
o.v. ^ = of value
Änderungen vorbehalten / Subject to change without notice
≤ ≤ ≤ ≤ ≤ 1%. The high degree of
max. 10mJ (^ = Capacitor 2µF,

Publicidad

Tabla de contenido
loading

Tabla de contenido