Important Health-related Information
• The ReliOn™ Premier Voice Blood Glucose Monitoring System
is intended for in vitro (outside the body) diagnostic use only
and for the quantitative measurement of glucose in capillary
whole blood.
• The ReliOn™ Premier Voice Blood Glucose Monitoring System
is intended for use by a single user only. Do not share the
meter and lancing device with another person.
• The ReliOn™ Premier Voice Blood Glucose Meter is designed
to minimize code-related errors in monitoring by using the
no-coding function.
• Glucose in blood samples reacts with the chemical in the
test strip to produce a small electrical current. The ReliOn™
Premier Voice Meter converts this electrical current to
glucose concentration.
• The ReliOn™ Premier Voice Blood Glucose Meter should be
used only with the ReliOn™ Premier Test Strips.
• An abnormally high or low red blood cell count (hematocrit
level over 60% or below 20%) may produce inaccurate results.
• Inaccurate results may occur in severely hypotensive (having
low blood pressure) individuals or patients in shock.
• Inaccurate low results may occur for individuals experiencing a
hyperglycemic-hyperosmolar state, with or without ketosis.
• Inaccurate results may occur in patients undergoing
oxygen therapy.
• Critically ill patients should not be tested with blood
glucose meters.