Precision: Precision studies were performed in a laboratory
using the ReliOn™ Premier Voice Blood Glucose Monitoring
System.
Within Run Precision
Blood avg.
Blood avg.
Blood avg.
Blood avg.
Blood avg.
Between Run Precision
Control avg.
Control avg.
Control avg.
This study shows that there could be variation of up to 4.1%.
52
38 mg/dL
86 mg/dL
125 mg/dL
189 mg/dL
334 mg/dL
35 mg/dL
116 mg/dL
324 mg/dL
SD = 1.9 mg/dL
SD = 3.2 mg/dL
CV = 4.1%
CV = 2.6%
CV = 2.8%
SD = 1.2 mg/dL
CV = 2.4%
CV = 2.3%