These tips are used for volumes between 2
l and 200 l. They are used with the PIP 20,
PIP 100, and the PIP 200 models.
Blue tips 1000
These tips are used for volumes between
100 l and 1000 l. They are used with the
PIP 1000 model.
White tips 5000
These tips are used for volumes between
1000 l and 5000 l. They are used with the
PIP 5000 model
7 RECOMMENDATIONS
The recommendations below will ensure
maximum accuracy and precision from your
pipettet.
Make sure to operate the pipette slowly
and smoothly
The depth of immersion in the sample
liquid should be the minimum necessary
and
should
aspiration.
The pipette should be held in a vertical
position.
Change the tip when volume setting is
changed or when a different liquid is to be
aspirated.
Change the tip if a droplet remains on the
end of the tip from the previous pipetting
operation.
Each new tip should be pre-rinsed with the
liquid to be pipetted.
Liquid should never enter the pipette
shaft. To prevent this :
Press and release the pushbutton slowly
and smoothly.
Never turn the pipette upside down.
Never lay the pipette on its side when
there is liquid in the tip.