pressure. The seal and the O-ring should
not be autoclaved.
The outside of the pipette can be cleaned
using a cloth dampened with isopropanol.
If there is an air bubble when the sample is
aspirated :
Eject the sample into its original vessel.
Check that the tip is properly immersed in
the same liquid.
Pipette the sample more slowly.
If the bubble appears for the second time,
replace the tip.
If a problem continues after carrying out the
above steps, send the pipette to your
representative.
Before returning a pipette to a represen-
tative, please ensure that the instrument is
completely free of any contamination
(chemical, microbian or radioactive).