Instant Flexible Digital Thermometer
EN
Owners Manual
WARNING
• Read instructions thoroughly before using digital thermometer.
• Choking hazard: Thermometer cap and battery may be fatal if swallowed. Do not allow children
to use this device without parental supervision.
• Do not use thermometer in ear. Designed use is for oral, rectal, and armpit (axilla) readings only.
• Do not place thermometer battery near extreme heat as it may explode.
• Note: Use of the probe cover may result in a 0.1°C (0.2°F) discrepency from actual temperature.
• It is recommended the performance should be checked every two years.
• Remove battery from the device when not in operation for a long time.
• The use of temperature readings for self-diagnosis is dangerous. Consult your doctor for the
interpretation of results. Self-diagnosis may lead to the worsening of existing disease conditions.
• Do not attempt measurements when the thermometer is wet as inaccurate readings may result.
• Do not bite the thermometer, doing so may lead to breakage and/or injury.
• Do not attempt to disassemble or repair the thermometer, doing so may result in inaccurate readings.
• After each use, disinfect the thermometer especially if device is used by more than one person.
• Do not force the thermometer into the rectum. Stop insertion and abort the measurement when
pain is present. Failure to do so may lead to injury.
• Do not use thermometer orally after rectal use.
• For children who are two years old or younger, please do not use the devices orally.
PLEASE READ CAREFULLY BEFORE USING
This digital thermometer provides a quick and highly accurate reading of an individual's body
temperature. To better understand its functions and to provide years of dependable results, please
read all instructions first.
This thermometer is manufactured according to the system of EN12470-3:2000+A1:2009,
ASTM E1112-00 (Reapproval 2011).
The manufacturer is EN ISO 9001:2008 and EN ISO 13485:2012/AC:2012.
CONTENT
1 Thermometer, 1 Owners Manual, 1 Storage Case
PRODUCT ILLUSTRATION
Figure 1
Probe
PRECAUTION
• The performance of the device may be reduced should one or more of the following occur:
· Operation outside the manufacturer's stated temperature and humidity range.
· Storage outside the manufacturer's stated temperature and humidity range.
· Mechanical shock (for example, dropping the device).
· Patient temperature is below room temperature.
• Portable and mobile RF communications can affect the device. The device needs special precau-
tions regarding EMC according to the EMC information pro vided in the accompanying documents.
SYMBOL EXPLANATION
Direct Current
Type BF Applied Part
!
Caution, consult accompanying documents
SPECIFICATION
Type
Range
Accuracy
Display
Memory
Battery
Battery life
Dimensions
Weight
Using enviroment
Storage & transp. condition Temperature: -20°C ~ 55°C (-4°F ~ 131°F) // Relative humidity: 15 % ~ 95 % RH
Ingress Protection Rating
Classificarion
Digital Thermometer (Not Predictive)
32.0°C-42.9°C (90.0°F-109.9°F) (°C/°F chosen by manufacturer)
± 0.1°C, 35.5°C-42.0°C (± 0.2°F, 95.9°F-107.6°F)
± 0.4°C under 35.5°C or over 42.0°C // (± 0.4°F under 95.9°F or over 107.6°F)
at ambient temperature of 18°C to 28°C (64.4°F to 82.4°F)
Liquid crystal Display, 3 1/2 digits
For storing the last measured value
One 1.5 V DC. button battery (size LR41 or SR41, UCC 392)
Approx. 200 hours
13.9cm x 2.3cm x 1.3cm (L x W x H)
Approx. 13 grams including battery
Temperature: 5°C ~ 40°C (41°F ~ 104°F) // Relative humidity: 15 % ~ 95 % RH
IP 27
Type BF
LCD
On/Off
LOT
Batch code
Manufacturer
European Authorized Representative
EC REP
Battery Cap
Button
!