Idexx Quanti-Tray Sealer PLUS Manual De Usuario página 8

Ocultar thumbs Ver también para Quanti-Tray Sealer PLUS:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Cleaning and Preventive Maintenance
Cleaning the Rubber Insert
You can autoclave the rubber insert or clean it with mild detergent, diluted bleach, or isopropyl alcohol, taking the usual precautions when
handling such liquids. Be sure to clean the rubber insert if it shows any signs of built-up grime and ensure the rubber insert is dry prior to use.
Cleaning the Outside of the Sealer
Clean the outside of the sealer with a soft, dry cloth or with a soft cloth moistened with water, mild detergent, diluted bleach, or isopropyl
alcohol, taking the usual precautions when handling such liquids.
Cleaning the Inside of the Sealer
Routine maintenance and prompt cleaning after a spill will help to maintain the performance of the sealer.
Only qualified personnel should open the access doors to clean the inside of the sealer.
CAUTION: Do not open the access doors of the sealer or tilt the sealer if sample is dripping from the unit. Wait until dripping stops and
the sealer is cool.
For troubleshooting or repairs other than cleaning, contact IDEXX Technical Service at +1 800 321 0207.
To clean the inside of the sealer:
1. Power off the sealer: Unplug the sealer, and allow it to cool for at least 30 minutes.
2. Open the sealer:
a) Use a standard flathead screwdriver to unlock the access panel, turning the screw 90° clockwise.
b) Press in the tabs on both sides of the lower access panel, and then slide the access panel out of the sealer and set aside.
c) Press the round indentation in the upper access door to push the door inward.
6
2a.
2b.
2c.

Publicidad

Tabla de contenido
loading

Productos relacionados para Idexx Quanti-Tray Sealer PLUS

Tabla de contenido