Copper With Tablet - Aqualytic AL100 Manual De Instrucciones

Ocultar thumbs Ver también para AL100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 214
Methods
GB
Cu
0.0.0
Zero
Test
Cu
RESULT
Zero
Test
Cu
RESULT
78

Copper with Tablet

0.3 – 5.0 mg/l Cu
a) free Copper
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Add one COPPER No. 1 tablet straight from the foil to the water
sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l free Copper.
b) total Copper
Add one COPPER No. 2 tablet straight from the foil to the same
water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l total Copper.
c) combined Copper
combined Copper = total Copper – free Copper
marks
marks
AL100_6a 05/2019

Publicidad

Tabla de contenido
loading

Tabla de contenido