solution and dried before plugging into the light source receptacle. Ensure the optical surface is clean
before engaging into the light source.
Caution
from the headlight module to the patient is not less than 12 inches (30.5 cm). If using less than 12
inches from the patient, the light intensity must be turned down.
Caution
unacceptable risk. If this determination is made arrangements should be taken to reduce the risk.
3. SPECIFICATIONS
Item
Lamp type
Power
Color temperature
Lamp life
Lamp replacement
Light guide adapter
Brightness control
Input voltage
Power consumption
Regulatory Approvals
Equipment Class
Mode of Operation
Water Resistant
Operating Environment
Temperature
Relative Humidity
Air Pressure
LIT034 Sunoptics Surgical
(English)
When using high intensity lightsources at full output, the recommended distance
User is responsible for determining if interruption of light output will create an
Specification
Cermax type 175 watt Xenon
175 Watt
5600º K
650 hours (typical)
Cartridge replacement from the side of the unit
Turret type to fit your choice of four: Storz, ACMI, Pentax, Wolf,
Olympus
Mechanical iris control from the front panel
100-120V AC, 50/60 Hz
220-240V AC, 50 Hz
250 watt max
UL60601, EN 60601-1, EN 60601-1-2, CAN/CSA, C22.2,
No 601.1-M90
BF-type
Continuous operation
Not Protected Equipment, IPX0
+10° to +40° C (50° to 104° F)
30 to 85%
700 to 1060 hPa
®
Rev. N
Date of Revision:07/02/14
Page 5 of 68