Cleaning And Disinfecting - Permobil ROHO Hybrid Elite Manual De Operación

Ocultar thumbs Ver también para ROHO Hybrid Elite:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 14

Cleaning and Disinfecting

Warnings:
- Cleaning and disinfecting are separate processes. Cleaning must precede
disinfection. Before use by a different individual: clean, disinfect, and check product
for proper functioning.
- Follow all manufacturer bleach container and germicidal disinfectant safety
guidelines.
- Note About Sterilization: High temperatures accelerate aging and will damage the
product assembly. The products featured in this manual are not packaged in a
sterile condition, nor are they intended or required to be sterilized prior to use. If
institution protocol demands sterilization: First, follow the cleaning and disinfecting
instructions; then, open the inflation valve and use the lowest possible sterilizing
temperature, but not to exceed 70°C (158°F), for the shortest time possible. DO NOT
autoclave.
Cover Care: Before cleaning and disinfecting, remove the cover from the cushion. Remove any
hook fastener pieces from the bottom of the cover.
To Clean the Cover: Machine wash in cold water (30°C) with mild
detergent, gentle cycle, or wipe clean with neutral detergent and
cold water. Tumble dry low.
To Disinfect the Cover: Hand wash in cold water (30°C) using 1
part household liquid bleach per 9 parts water. Keep the product
wet with the bleach solution for 10 minutes, and then rinse
thoroughly. Or machine wash in hot water (60°C). Tumble dry low.
Cushion Care Cautions: DO NOT machine wash or machine dry the cushion insert or
the contoured foam base. • Use of the following may cause damage to the cushion:
abrasives (steel wool, scouring pads); caustic, automatic dishwasher detergents; cleaning
products that contain petroleum or organic solvents, including acetone, toluene, Methyl Ethyl
Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover; oil-based lotions, lanolin, or
phenolic-based disinfectants; ozone gas; ultraviolet light cleaning methods.
Warning: DO NOT allow water or cleaning solution to enter the cushion. Make
sure the inflation valve is closed. Make sure the Smart Check is detached from the
cushion. Water in a cushion with Sensor Ready (SR) technology can result in damage to
the Smart Check and cause a false reading when using Smart Check.
To clean and disinfect the Cushion Insert: Carefully remove the cushion insert from the
contoured foam base. Close the inflation valve(s). Hand wash, using mild liquid
hand soap, hand dishwashing liquid, laundry detergent, or a multipurpose
detergent (Follow product label instructions.) Gently scrub all cushion surfaces,
using a soft plastic bristle brush, a sponge or a cloth. Rinse with clean water. Air dry. DO NOT
expose to direct sunlight. To disinfect, repeat cleaning instructions, using 1 part household
liquid bleach per 9 parts water. Keep the product wet with the bleach solution for 10 minutes.
Rinse thoroughly with clean water. Air dry.
Caution: Rinse thoroughly. Detergent residue may cause the air cells to stick together.
ROHO Hybrid Elite Cushion Operation Manual
8
30C
+
or
30C
1:9

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido