COMMENTS ON VERIFICATION

Do not take this section into consideration, if
the weighing instrument does not bear
"green M" sticker.
Measuring
instruments
are
reverified according to the legal
requirements applicable in the
individual Member States. See
data plate for the year (near the
CE marking) the measuring instrument
is installed and put into service. This
date authoritative in two years but
Period of circulation can be changed in
your
country.
End
of
period
of
circulation you must go to relevant
ministry with your machine.
If the instrument connected to additional
equipment, the "red M" sticker will be
affixed to the additional equipment, thus
permitting it to be put into service and
trade with the exception of direct sales
to the public.
After
the
periodic
or
repairs,
for
inspection
and
sealing
process
authorized should take empty counter
"Service Tag" from your services. Then,
with this label should go to inspection
and should be determined. Details
about the inspection process can be
found
on
the
website
from
www.temterazi.com
Important stickers that permitting the
instrument to be put into service and
used in trade are;
Data plate with "green M"
sticker
Security sticker(s)
If modifications of any kind are made to
stickers that given above after the
weighing instrument being put into
service, the competent Weights and
Measures Authorities must be notified.
Detail Security of Seals:
The count of event counter is displayed
during start-up for two seconds and is
identified as C XXXX where XXXX is the
count. This count must be the
same with inscribed count of
event counter
on a label is
placed near the data plate.
If this two numbers are
not same,it means that the seal
breaks down.
Count of event counter that
Event Counter
displayed during start-up
o
The
event counter is incremented
each time the calibration or any
metrological configuration parameters
are changed. The security of weighing
instrument will be broken. In this case
or any damage on event counter
sticker
or
data
plate,
weighing
instrument has verification done again.
The responsibility of using weighing
instrument with out verification belongs
to user.
- 5 -
o
NOTE: Calibration and Metrological
configurations are secured with a
special password. Therefore these
settings can not changed mistakenly.
The increasing of event counter is an
argument of consciously arrangement
The responsibility of using weighing
instrument in this situation belongs to
user.
Places of Verification Stickers:
o
There must be a security sticker with
TEM logo(Hologram) on data plate
Data plate of weighing instrument
There must be a security sticker with
o
TEM logo(Hologram) on Event Counter
Sticker. On indicators the load cell
serial number does not exists on
Event Counter
Service
Counter
Sticker
Sticker
In the weighing instrument, there must be an
security sticker with TEM Logo(Hologram) to
protect measurement components.
o
If there is platform connection to indicator;
o
Load cell connector bears the serial
number(B) of weighing instrument on
its identification label(A). These serial
numbers must be same.
o
Load receptor bears the seri al
number
of
the
weighi ng
instrument on its identificati on
label.
- 6 -