Problem
The connector on the
end of the power cable is
plugged into the socket
on the rear of the device
properly.
A faint whistle can be
heard coming from the
ULTRASURGERY
handpiece during
operation.
The device is switched
on but does
not work, the message
WARN appears on the
display.
The device is switched
on but will not work, the
message WARN appears
on the display.
Possible cause
The connector of the
footswitch is not properly
plugged into the socket.
The footswitch will not
work.
The tip is not correctly
tightened onto the
handpiece.
The tip is not fitted correctly
into the handpiece.
The tip is worn, broken or
deformed.
The connector of the cord
is wet.
Cord not connected to the
device.
Lack of continuity of a lead
in the cord.
Handpiece failure.
Malfunctioning of the
tuning circuit.
26
solution
Insert the footswitch
connector
properly.
Contact the nearest
dealer or authorized
Woodpecker service
centre.
Unscrew the tip and
screw it back into place
correctly.
Unscrew the tip and
screw it back into place
correctly.
Replace the tip.
Dry the connectors.
Connect the cord to the
device.
Contact the nearest
dealer or authorized
Woodpecker service
centre.
Contact the nearest
dealer or authorized
Woodpecker service
centre.
Contact the nearest
dealer or authorized
Woodpecker service
centre.