Specifications
Measuring range
Measuring site
Reference body site
Operation mode
Laboratory accuracy
Display resolution
Measure time
Operating temperature
range
Storage and transport
temperature range
Clinical accuracy
Dimension
Weight
Battery
Battery life
Ingress protecting rating
www.self-diagnostics.com
32,0°C ~ 43,0°C (89,6°F ~ 109,4°F)
Ear canal
Oral (This thermometer converts the ear tempe-
rature to display its "oral equivalent.")
Adjusted mode
± 0.2°C (0.4°F) during 35.0°C ~ 42.0°C (95.0°F ~
107.6°F) at 15°C ~ 35°C (59.0°F ~ 95.0°F)
operating temperature range ± 0.3°C (0.5°F) for
other measuring and operating temperature range
0.1°C or 0.1°F
Approximately one second
10°C ~ 40°C (50°F ~ 104°F), 15% ~ 85% RH,
non-condensing
-25°C ~ 55°C (-13°F ~ 131°F), 15% ~ 95% RH,
non-condensing
Clinical bias: -0.03°C (-0.5°F)
Clinical repeatability: 0.19°C (0.4°F)
Limits of agreement: 0.5°C (0.9°F)
106 x 33 x 47 mm
Approx. 34 grams (with battery)
One CR2032 lithium cell battery
Approx. 1 year / 3000 readings
IP22
EN
17