Care And Cleaning
1. Probe covers are intended for single use only!
Absent, damaged, perforated, or soiled probe covers can lead to
inaccurate temperature readings.
2. The probe window must be kept clean, dry and undamaged at all
times to ensure accurate readings. The accuracy of temperature
readings can be affected by damage to the probe window, or the
presence of dirt and ear wax on the probe window.
3. Fingerprints, earwax, dust and other soiling compounds reduce
transparency of the window and result in lower temperature read-
ings. If the thermometer is accidentally placed in the ear without
a probe cover installed, the window must be cleaned immediately.
4. The probe window is the most delicate part of the thermometer.
To safely clean the window, gently wipe its surface with a cotton
swab slightly moistened with isopropyl alcohol and immediately
wipe dry with a clean cotton swab. After cleaning, allow at least
5 minutes drying time before taking temperatures.
Note: Do not use any chemical other than isopropyl alcohol to
clean the probe window.
5. Use a soft, dry cloth to clean the thermometer display and exterior.
6. The thermometer is not waterproof. Do not submerge the unit in
water when cleaning.
14
Infrared Ear Thermometer