VI- Positioning the tubes

VII - Malfunctions list
Problem
The pump is not working
properly
Commands are not
working
Irrigation flow too low
Tube lifetime to low
Beats into the tube
Abnormal noise coming
out of the pump
NT010
K – 2015/11
Incorrect
How to position the tubes: front and top views
Possible reason
No electrical power
The pump wheel stalls
No control of the irrigation
No control of the suction
(CM201)
Bad setting of the flow rate
Tube partially or totally clogged
Wearing of the tube
Tube/pump incompatibility
Tube/pump incompatibility
Bad setting up of the tube into
the pump block
Use generating important beats,
due to the pump, the speed of
the pump, the pression in the
tube
Deterioration of the device
Correct
Solution
Verify that power supply switch is on
Verify the power cord connection
Verify the tube connection
Verify that the pneumatic petal is switched on
Verify that the handle is switched on
Regulate the flow using the potentiometer
Set up a new sterile tube, Integra MicroFrance
MCEN84S
Set up a new sterile tube, Integra MicroFrance
MCEN84S
Set up a new sterile tube, Integra MicroFrance
MCEN84S
Set up a new sterile tube, Integra MicroFrance
MCEN84S
Set up a new sterile tube, Integra MicroFrance
MCEN84S
Contact Integra MicroFrance
Contact Integra MicroFrance