Electromagnetic Emissions And Immunity - Integra MicroFrance MCEN80 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 46
XVII - Electromagnetic emissions and immunity
Origin of
Frequency
interference
range
0.15 to 0.5 MHz
AC supply
0.5 to 30 MHz
30 to 230 MHz
Casing
230 MHz to
1 GHz
Harmonic
0 kHz to 2 kHz
currents
Flicker
-
Origin of
interference
Electromagnetic
field with amplitude
modulation radio
frequency
Electrostatic
discharges
Magnetic field in
power grid
frequency
NT010
K – 2015/11
Limits
Quasi-peak:
79 dBµV
Average
66 dBµV
Quasi-peak
73 dBµV
Average
60 dBµV
30 dBµV/m
at 10 m
37 dBµV/m
at 10 m
Class A
Pst, Plt
Dc, D(500),
DMAX
Casing access
Test specifications
80 to 2500 MHz
3 V/m
AM 80%
1 kHz
+/- 6 kV in contact
+/- 8 kV in air
50 Hz
3 A/m efficient
Emission
Basic standard
CISPR 11
Class A
Satisfactory
Group 1
CISPR 11
Satisfactory
Class B
CEI 61000-3-2
Satisfactory
CEI 61000-3-3
Satisfactory
Immunity
Suitability
Basic standard
criteria
See
CEI 61000-4-3
standard
See
CEI 61000-4-2
standard
See
CEI 61000-4-8
standard
Results
Comments
The statistical
evaluation of the
standard is applicable.
The statistical
evaluation of the
standard is applicable.
Nothing to report.
Considering the low
power consumption,
the pump satisfies the
limits without additional
measurements.
Results
The specified
level is that of the
carrier, with
Satisfactory
efficacy value,
before
modulation.
See the basic
standard for the
applicability of
Satisfactory
discharges in
contact and/or in
the air.
Satisfactory
-
Page 44 sur 91
Comments