WARNING: Never point a loaded serter toward any body part where insertion is not desired.
An accidental button-push may cause the needle to inject the sensor in an undesired location,
causing minor injury.
10a
10b
10a. Hold the serter
steady against
your cleaned
insertion site,
without pushing
the serter too
deep into your
skin.
Note: Failing to hold the serter securely flat against your body may allow the serter to spring back after
pressing the buttons, and result in improper insertion of the sensor.
Note: The arrow on the side of the serter aligns with the needle inside the
serter.
10c
10b-10c. Press and
release the bump
on both buttons
at the same time,
while holding the
serter flat against
your body.
10d
10d. Continue holding
the serter flat
against your
body to allow the
adhesive to stick
to your skin.
- 7 -
10e
10e. Slowly lift the
serter away from
your body, making
sure that the
buttons are not
pressed.