Product description:
V-Print SG is a light-curing acrylic for the generative production of biocompatible dental
drilling templates. Drilling templates made of V-Print SG can be steam sterilised at 134 °C for
max. 5 minutes.
Indications:
– Dental drilling templates
Safety and precautions:
Avoid direct contact between the skin and liquid or uncured V-Print SG. The wearing of protective
clothing is recommended. This should encompass gloves (nitrile), safety goggles and a long-sleeved
lab coat. Further information on the handling of the product can be found on the safety data sheet.
V-Print SG can trigger sensitisation or irritation of the skin and/or respiratory organs. Repeated and/
or prolonged exposure through contact and/or inhalation, including of dusts, can trigger irritations
or even result in serious illness.
Note: Use separate material containers and cleaning baths for each printing material so as to rule out
the possibility of cross-contamination.
Preparation:
V-Print SG is a light-curing acrylic designed for the generative production of articles using a
385 nm LED DLP printer. Information on approved printers is available from the corresponding
device manufacturers and from VOCO.
The following construction conditions must be observed for indication-appropriate CAD construc-
tions:
– Minimum wall thickness 3 mm
Prepare a print job using slicing software. In addition to the material-dependent construction specifi-
cations in these instructions for use, please also observe the dependencies on the positioning, type
of support and adaptation from our other documents for your construction.
V-Print SG can be printed in the following layer thicknesses:
– 50 μm and 100 μm
Processing:
The material container should be filled directly before printing is started. When doing so, ensure the
material is as free from bubbles as possible and pay attention to the fill level.
Note: Do not shake the material before starting printing.
Start the print job observing the previously selected parameters.
We recommend allowing a dripping time of approx. 10 minutes once printing is complete. Then
detach the printed objects carefully from the build platform.
Following this, the printed objects must be cleaned, dried and post-exposed in order to guarantee
the requisite product characteristics. A detailed description of the two steps outlined above can be
found under Post-processing. Post-process the printed object without delay so as to avoid distor-
tion of its shape and ensure the required precision.
Remove the excess material from the material container following printing, so as to allow you to clean
and check the container. The print material can be returned to its original packaging. If necessary, a
separate container can also be used (HDPE, not light-transmissive, airtight).
Please ensure that the excess material intended for further use is free from contaminants and po-
lymerised components. To this end, a stainless steel sieve or similar instrument should always be
used when transferring material.
Post-processing:
Cleaning:
We recommend a solvent-resistant cleaning mat for cleaning.
Cleaning should be performed with isopropanol (≥ 98% purity) in an unheated ultrasound bath. The
respective operating manuals and/or instructions for use provided by the manufacturers must be
observed.
The printed objects must be cleaned in two or optionally in three stages.
Pre-cleaning:
– Optional: Pre-clean printed objects carefully for approx. 15 seconds by repeatedly immersing them
in a beaker of isopropanol. When doing so, hold the component by a support using tweezers.
– Pre-clean the printed objects for 3 minutes in a reusable isopropanol ultrasound bath. Position
uncleaned printed objects in the ultrasound bath in such a way that any openings face downwards.
Note: The bath's cleaning efficacy decreases the more it is used. Replace the bath when the
cleaning efficacy decreases.
Main cleaning:
Clean the printed objects for a further 2 minutes in a fresh isopropanol ultrasound bath.
Then dry them carefully with compressed air.
If there are any resin residues on the printed object after the main cleaning or any exits from the
undercuts during drying, the printed object can be immersed in the main cleaning bath again
briefly before being dried again.
Preparation for post-exposure:
Prior to post-exposure, remove the support structures carefully using a rotary instrument (e.g. a
cutting disc) and without exerting pressure, taking care to make the cut as close as possible to the
printed object. Use a suction device. Remove any remaining resin dust carefully using compressed
air. Then rinse printed objects for a few seconds with fresh isopropanol and remove any dust using
a fine, clean brush if necessary. Then dry the printed objects again carefully using compressed air.
Bonding of drill sleeves:
Coat the exterior of the sleeve and the intended hole in the drilling template with a thin
layer of liquid V-Print SG prior to the post-exposure. Insert the sleeves and remove any excess
material to ensure marginal integrity. Then post-expose the drilling template as described under
"Post-exposure".
Post-exposure:
No protective gas atmosphere is required. However, ensure that the printed objects do not cast
shadows over or touch each other, as this could result in incomplete post-curing.
The post-exposure can be performed with the following devices:
– Otoflash G171 xenon photoflash unit
– LC-3DPrint Box
UV light box
1
In printed objects post-exposed with the Otoflash G171, a cooling-off period of at least 2 minutes
with the lid open must be included after the first 2,000 flashes.
2 × 2,000 flashes
30 minutes
Once the cooling-off period has elapsed, the printed objects should be turned and light-cured with
a further 2,000 flashes.
Insufficient cooling can distort the shape of the printed object.
Finishing:
As a general rule, work with a low contact pressure and speed. This guarantees constant results and
also reduces the risk of undesirable signs of processing.
Use a fine-toothed carbide bur or similar to sand the support stubs. The same bur can also be used
for subsequent finishing of special structures.
For precise sanding, e.g. between the support strut and the printed object, it is recommended that
the surface be processed in the corresponding areas with sandpaper, if necessary with different grain
sizes. A corresponding result can also be achieved with coarser or finer silicone polishers.
Final cleaning:
Clean the object thoroughly. Begin by removing large residues with the steam jet. The subsequent
final cleaning can be performed via brief storage in an unheated water ultrasound bath.
Sterilisation:
The sterilisation process is validated according to DIN EN ISO 17664 and can be used with the
equipment that complies within EN 13060. The instructions for use from the manufacturer have to be
observed and followed. The user is responsible for the sterility of the product. Only use completely post-
exposed and cooled drilling templates to undergo the steam sterilisation! Otherwise a deformation
of the printed object cannot be excluded.
Technical data of the sterilisation process:
134 °C
2,07 - 2,17 bar
Disinfection:
After use, drilling templates produced from V-Print SG can be disinfected with alcohol- or
aldehyde-based disinfectant solutions (e.g., ethanol, MD 520 from Dürr, Cavex Impre Safe from
Cavex). Observe the manufacturer's instructions for use.
Information, precautionary measures:
V-Print SG may only be used intraorally in its completely cured state. Pay attention to the finishing
process.
V-Print SG contains methacrylates. V-Print SG should not be used in case of known hypersensitiv-
ities (allergies) to any of these ingredients.
Storage:
Store at between 15 °C and 28 °C. Seal bottle again immediately after use. The material will cure if
exposed to light. Do not use after expiry date.
Our preparations have been developed for use in dentistry. As far as the application of the prod-
ucts delivered by us is concerned, our verbal and/or written information has been given to the best
of our knowledge and without obligation. Our information and/or advice do not relieve you from ex-
amining the materials delivered by us as to their suitability for the intended purposes of application.
As the application of our preparations is beyond our control, the user is fully responsible for the
application. Of course, we guarantee the quality of our preparations in accordance with the
existing standards and corresponding to the conditions as stipulated in our general terms of sale
and delivery.
This material has been developed solely for use in dentistry. Processing should be done strictly
according to the instructions for use. VOCO recognizes its responsibility to replace products if
proven to be defective. VOCO does not accept liability for any damage or loss, directly or indirectly,
stemming from the use of or inability to use the products described. Before using, it is the responsibility
of the user to determine the suitability of the product for its intended use. The user assumes all risk and
liability in connection therewith. Descriptions and data constitute no warranty of attributes and are
not binding.
CAUTION: U.S. Federal Laws restrict this device to sale by or on the order of a dentist.
No person is authorized to provide any information which deviates from the information provided
in the instructions for use.
For questions or comments, please call 1-888-658-2584.
Keep this material out of reach of children.
For dental use only.
An explanation of the symbols used in labeling can be found at
www.voco.com/us/symbols
1
Or a similar device. Please consult your device's manufacturer for further information.
DE
Gebrauchsanweisung
Produktbeschreibung:
V-Print SG ist ein lichthärtender Kunststoff zur generativen Herstellung von dentalen biokom-
patiblen Bohrschablonen. Aus V-Print SG hergestellte Bohrschablonen sind bei 134 °C für max.
5 Minuten dampfsterilisierbar.
Indikationen:
– Dentale Bohrschablonen
Sicherheit und Vorsorge:
Der direkte Hautkontakt mit flüssigem oder nicht nachpolymerisierten V-Print SG ist zu vermeiden.
Das Tragen von Schutzkleidung wird empfohlen. Eine etwaige Schutzausrüstung sollte Handschuhe
(Nitril), Schutzbrille und langärmligen Kittel umfassen. Weitere Informationen über die Handhabung
kann dem Sicherheitsdatenblatt entnommen werden.
V-Print SG kann zu einer Sensibilisierung oder Reizung der Haut und/oder Atmungsorgane führen.
Wiederholter und/oder längere Exposition durch Kontakt und/oder Einatmung, auch von Stäuben,
kann zu Reizungen oder bis hin zu schwerwiegenderen Erkrankungen führen.
Hinweis: Verwenden Sie für jedes Druckmaterial separate Materialwannen und Reinigungsbäder, um
Kreuzkontaminationen auszuschließen.
Vorbereitung:
V-Print SG ist ein lichthärtender Kunststoff, konzipiert für die generative Fertigung mittels 385 nm
LED-DLP-Drucker. Informationen zu Druckerfreigaben können bei entsprechenden Geräteherstellern
oder bei VOCO angefragt werden.
Für eine indikationsgerechte CAD-Konstruktion sind folgende Konstruktionsbedingungen zu berück-
sichtigen:
– Mindestwandstärke 3 mm
Bereiten Sie einen Druckjob mittels Slicing-Software vor. Beachten Sie neben den materialbeding-
ten Konstruktionsvorgaben aus der vorliegenden Gebrauchsinformation, auch die Abhängigkeiten
von Positionierung, Support-Art und Passung aus unseren weiterführenden Unterlagen für Ihre
Konstruktion.
V-Print SG kann in folgenden Schichtstärken gedruckt werden:
– 50 µm und 100 µm
Verarbeitung:
Das Befüllen der Materialwanne sollte unmittelbar vor Druckbeginn erfolgen. Es ist darauf zu achten,
dass das Material möglichst blasenfrei unter Beachtung des Füllstandes einzufüllen ist.
Hinweis: Das Material vor Druckbeginn nicht schütteln.
Starten Sie den Druckauftrag unter Berücksichtigung der von Ihnen zuvor gewählten Parameter.
max. 5 min
1

Publicidad

loading

Resumen de contenidos para VOCO V-Print SG

  • Página 1 Note: Use separate material containers and cleaning baths for each printing material so as to rule out Information, precautionary measures: the possibility of cross-contamination. V-Print SG may only be used intraorally in its completely cured state. Pay attention to the finishing process. Preparation: V-Print SG contains methacrylates.
  • Página 2 Avant la post-exposition, appliquer au pinceau sur l‘extérieur de la douille et sur l‘évidement V-Print SG enthält Methacrylate. Bei bekannten Überempfindlichkeiten (Allergien) gegen die prévu dans le guide de forage une fine couche de V-Print SG liquide. Insérer la douille et retirer les Inhaltsstoffe von V-Print SG ist auf die Anwendung zu verzichten.
  • Página 3 Antes de la post-exposición, recubra ligeramente la cara exterior del casquillo y la entalladura pre- post-traitement. vista en la plantilla de perforación con V-Print SG líquido. Coloque el casquillo y elimine el material excedente del sellado marginal. A continuación, post-exponga la plantilla de perforación tal y como V-Print SG contient des méthacrylates.
  • Página 4 La procedura di sterilizzazione è validata dalla norma DIN EN ISO 17664 e può essere utilizzata con Evitare il contatto diretto della pelle con V-Print SG in forma liquida o non post-polimerizzata. Si con- un dispositivo conforme alla norma EN 13060. È necessario osservare e seguire le istruzioni d’uso siglia l’uso di indumenti protettivi, che comprendono guanti (in nitrile), occhiali di protezione e ca-...