10.9. Información con Respecto a la Compatibilidad electromagnética
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The VOLCANO MEDIC Vaporization System is intended for use in the electro-
magnetic environment specified below. The customer or the user of the VOL-
CANO MEDIC Vaporization System should assure that it is used in such an en-
vironment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
flicker emissions
IEC 61000-3-3

Fabricante

Storz & Bickel GmbH & Co. KG
In Grubenäcker 5-9,
78532 Tuttlingen, Alemania
Electromagnetic environment - guidance
The VOLCANO MEDIC Vaporization
System uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The VOLCANO MEDIC Vaporization
System is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
50
10. EL GENER ADOR DE AIRE CALIENTE
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The VOLCANO MEDIC Vaporization System is intended for use in the electro-
magnetic environment specified below. The customer or the user of the VOL-
CANO MEDIC Vaporization System should assure that it is used in such an en-
vironment.
Immunity test
IEC 60601
test level
Electrostatic
± 6 kV contact
discharge (ESD)
discharge
IEC 61000-4-2
± 8 kV air
discharge
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV line(s) to
IEC 61000-4-5
line(s)
± 2 kV line(s) to
earth
Compliance
Electromagnetic
level
environment – guidance
± 6 kV contact
Floors should be wood,
discharge
concrete or ceramic tile.
If floors are covered with
± 8 kV air
synthetic material, the
discharge
relative humidity should
be at least 30 %.
± 2 kV for power
Mains power quality
supply lines
should be that of a
typical residential,
not applicable:
commercial or hospital
input/output lines
environment.
are non-existent
± 1 kV line(s) to
Mains power quality
line(s)
should be that of a
typical residential,
± 2 kV line(s) to
commercial or hospital
earth
environment.