10. EL GENER ADOR DE AIRE CALIENTE
Immunity test
IEC 60601
test level
Voltage dips,
<5 % Uτ
short
(>95 % dip in Uτ)
interruptions and
for ½ cycle
voltage
variations on
40 % Uτ
power supply
(60 % dip in Uτ)
input lines
for 5 cycles
IEC 61000-4-11
70 % Uτ
(30 % dip in Uτ)
for 25 cycles
<5 % Uτ
(>95 % dip in Uτ)
for 5 s
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: Uτ is the a.c. mains voltage prior to application of the test level.
Compliance
Electromagnetic
level
environment – guidance
<5 % Uτ
Mains power quality
(>95 % dip in Uτ)
should be that of a typical
for ½ cycle
residential, commercial or
hospital environment. If
40 % Uτ
the user of the VOLCANO
(60 % dip in Uτ)
MEDIC Vaporization
for 5 cycles
System requires
continued operation
70 % Uτ
during power mains
(30 % dip in Uτ)
interruptions, it is
for 25 cycles
recommended that the
VOLCANO MEDIC
<5 % Uτ
Vaporization System be
(>95 % dip in Uτ)
powered from an
for 5 s
uninterruptible power
supply.
3 A/m
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical residential,
commercial or hospital
environment.
52
10. EL GENER ADOR DE AIRE CALIENTE
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The VOLCANO MEDIC Vaporization System is intended for use in the electro-
magnetic environment specified below. The customer or the user of the VOL-
CANO MEDIC Vaporization System should assure that it is used in such an en-
vironment.
Immunity test
IEC 60601
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5
GHz
Compliance
Electromagnetic
level
environment – guidance
Portable and mobile RF
communications
equipment should be
used no closer to any part
of the VOLCANO MEDIC
Vaporization System,
including cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended
separation distance :
d = 1,2 √P
3 Vrms
d = 1,2 √P
3 V/m
für 80 MHz bis 800 MHz
d = 2,3 √P
für 800 MHz bis 2,5 GHz
53