SEITE 7 - 148 x 210 mm - 3100021789/01 - SCHWARZ - 20-053 (kn)
4 LEDs = 90%, 3 LEDs = 80%, 2 LEDs = 70%,
1 LED = 60%.
If the light intensity is below 100% (fewer than 5 LEDs
are ON), check the light guide for contamination or
defects.
Either: clean the light guide of any contamination.
Or: clean the glare shield on the handpiece; see
"Care" for instructions about both procedures.
Or: replace the defective light guide with a new light
guide. Or: if the measures described above do not
result in any improvement, call 3M Customer Service
or the appropriate dealer.
Recommended Cure Techniques with 3M Adhesi-
ves
• Rotate the light guide into the desired position for
polymerization.
• A barrier sleeve may be used. Note that sleeve can
impact power and irradiance of a curing light. It is
critical to ensure the sleeve is powder free and the
seam is not over the tip of the light guide and that the
sleeve fits well. Intermediate level disinfection (ma-
nual disinfection with ready-to-use disinfectant wipes)
is indicated when a sleeve is used, otherwise the light
guide should undergo high level disinfection (manual
disinfection using disinfectant bath, see "Reproces-
sing the Light Guide").
The following sleeves have been tested:
Sleeve
®
Cure Sleeve
Tip Covers (short) – Kerr
Complete Curing Light Sleeves-Demi –
Kerr
TIDIShield™ Custom Fit Curing Light
Sleeves – TIDI Products
®
SmartLite
Max Lens Cover Sleeve –
Dentsply
Curelastic™ Cure-Light Sleeve
(Steri-Shield)
Sleep Mode
Once the handpiece is placed in the charger, all internal
functions and LEDs are automatically turned off as the
handpiece switches to sleep mode. This reduces the
power consumption of the battery to a minimal level.
Outside the charger, the handpiece is also switched to
sleep mode, if it is not used for approx. 5 min.
To terminate the sleep mode press the START button.
- The sleep mode termination signal (two short
audible signals) is emitted indicating that the
handpiece is ready for operation; the handpiece
displays the latest selected exposure mode and
time settings.
Acoustical Signals - Handpiece
An acoustical signal is emitted
• every time a button is pressed,
• every time the light is turned ON,
• 1 time after 5 sec exposure time, 2 times after 10 sec,
3 times after 15 sec.
Exception: in continuous mode; an acoustical signal
is emitted every 10 seconds.
Two acoustical signals are emitted
• every time the sleep mode is terminated by pressing
the START button,
• every time the light is turned OFF.
A 2-sec error signal is emitted, if
Impacts
Power or
• the handpiece over-heats,
Irradiance
• the battery lacks sufficient charge.
The acoustical signals from the handpiece can be turned
No
off (except for the 2-sec error signal). Follow these
instructions to turn them off. Put the handpiece in sleep
No
mode, e.g., by setting it in the charger. Take the device
from the charger; press first the TIME button, then the
No
START button. This changes the sleep mode of the
hand piece back to active and the status "Acoustical
signals activated" to "Acoustical signals deactivated".
No
The acoustical signals can be reactivated by following
the same procedure.
Yes
7