SEITE 4 - 148 x 210 mm - 3100021789/01 - SCHWARZ - 20-053 (kn)
Light intensity (between
400 and 515 nm):
1470 mW/cm
(independent of battery power
level)
Light emission area:
60-65 mm
Intermittent operation: The device has been designed
solely for short-term operation.
Typical operating time at room
temperature (23° C): 7 min,
at 40° C ambient temperature:
1 min on, 15 min off (cooling-off
period)
Total exposure time with new,
fully charged battery:
typically 120 min
Dimensions:
Diameter: 28 mm
Length:
Weight:
250 g (incl. light guide)
Typ B
Charger and Handpiece
Time to charge empty battery: approx.1.5 h
Operating temperature:
Relative humidity:
Atmospheric pressure:
Total height with handpiece
inserted in the charger:
Year of manufacture:
Transport and Storage Conditions:
Ambient temperature:
-20° C up to +40° C /
-4° F up to +104° F
–20°C
–4°F
Relative humidity:
30% up to 75%
30%
Atmospheric pressure: 700 hPa up to 1060 hPa
700
Subject to technical modification without prior notice.
4
2
-10%/+20%
2
(optically active)
270 mm
10° C up to 40° C /
59° F up to 104° F
30% up to 75%
700 hPa up to 1060 hPa
180 mm
see rating plate
+40°C
+104°F
75%
max.
1060
hPa
Installation of the Unit
Factory Settings
The factory settings of the unit are as follows:
• 10 sec exposure time
Initial Steps
Charger
Please ensure first that the voltage stated on the
rating plate corresponds to the existing main supply
voltage. The rating plate is attached to the bottom
of the unit.
Place the charger on a level surface.
Connect the power cable of the charger to the power
supply.
- The green LED on the charger showing that the
unit is ready for operation comes on. This shows
that the unit is ready for operation; please refer
to the section "Operating Status Display of the
Charger".
Light Guide/Handpiece
Never place the handpiece in the charger without the
battery inserted in the handpiece!
Place the glare shield on the front of the unit.
Autoclave the light guide prior to first use.
Then attach the light guide firmly to the handpiece.
Inserting the Battery
Remove the protective cap from the battery and keep
with the unit packaging.
Slowly insert the battery with the thread side towards
the metal housing into the handpiece until it stops.
Screw the battery clockwise by hand until the seal is
pressed firmly against the metal housing. The unit
will not function properly if the battery is not screwed
in completely!