The following parameters can be changed in the set-up menu: Default settings are shown bold.
Set-up parameter:
TOF s flashing (TOF s repetition time)
TOF s repetition time can be adjusted between 1 and
I
3
:
60 minutes.
Surface electrode + stimulation flashing (stimulation units)
mA:
Surface electrode stimulation strength in milli-ampères.
µC:
Surface electrode stimulation strength in micro-coulomb.
Surface electrode + stimulation flashing (pulse width)
200 µs:
Default surface stimulation pulse width can be selected
to 200 or 300µs.
Surface electrode + stimulation flashing (stimulation size)
50 mA:
Default surface electrode stimulation strength can be
adjusted between 0 and 60 mA (0 and 12/18 µC).
Needle electrode + stimulation flashing (stimulation units)
µC:
Needle stimulation strength shown in micro-coulomb.
mA: Needle stimulation strength shown in milli-ampères.
Needle electrode + stimulation flashing (stimulation size)
0.0 µC:
Default needle electrode stimulation strength can be
adjusted between 0.0 and 6.0 mA/0.24 µC.
P flashing (P button function)
- - :
no function assigned to P button
DBS 3.3:
DBS 3.3 stimulation assigned to P button
DBS 3.2:
DBS 3.2 stimulation assigned to P button
50 Hz:
50 Hz tetanic stimulation assigned to P button
100 Hz:
100 Hz tetanic stimulation assigned to P button
Lo + loudspeaker + transducer flashing (Low TOF alarm)
- - :
Low TOF alarm not active
1-4:
Low TOF alarm will sound if the number of TOF
responses are below the number of set responses.
3 - 100%:
Low TOF alarm will sound if the TOF ratio is below
set percentage.
17 | TOF-Watch SX Operator manual 33.512/A
Set-up Display