Technical specifications
Microprocessor-controlled.
Custom design LCD with numerical presentation of twitch data.
Automatic calibration:
Stimulation modes:
Output (accuracy ±5% of full scale value):
Surface electrodes
Needle electrodes
Acceleration transducer:
Temperature sensor:
Battery:
Supply voltage:
Type:
Weight:
Dimensions:
Safety:
EMC:
11
Symbols
Attention, consult accompanying documents.
!
Type BF equipment.
IPXO
Degree of protection against ingress of water (non-protected).
Complies with Medical Device Directive, 93/42/EEC (class 2A).
0543
22 | TOF-Watch SX Operator manual 33.512/A
Automatic adjustment for optimal gain and 100%
reference.
TOF (Train Of Four)
TOF s (Train Of Four) with programmable repetition time.
PTC (Post Tetanic Count)
1 Hz twitch
0.1 Hz twitch
DBS3.3 and 3.2 (Double Burst)
Tetanic stimulation (Burst), 5 s - 50 Hz or 100 Hz
Constant current, 0 - 60 mA (0 - 12/18 µC) up to 5 kOhm.
Monophasic, 200 µs or 300 µs pulse width.
Constant current, 0 - 6 mA or (0 - 0.24 µC) up to 5 kOhm.
Monophasic, 40 µs pulse width.
Acceleration transducer (accuracy ±5% of full scale value).
Range 20.0 - 41.5 °C (accuracy ±0.5 °C)
9V Alkaline or NiCd (6LR61 / 6AM6).
9V=, 220 mA max.
BF
Approximately 250 gram with battery.
200 x 90 x 50 mm.
Complies with IEC601-1.
Complies with EMC directive, 89/336/EEC, § 10 part. 1
when used with standard or optional accessories.