Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

SOFTWARE
S YSTE M R E QUIRE MENTS
software is designed to run under Windows. Suitable versions are Windows
2000, Windows XP, Windows Vista and Windows 7.
The recommended minimum computer
Intel Pentium processor
VGA display with 640x480 resolution
USB 2.0 port
INSTALLATION
1. Insert the disk provided
2. Installer should run automatically (if not, select the appropriate drive and double click
on Setup.exe)
3. Follow the on-screen instructions
NOTE Do not connect the PCE-IR 52
installed
C ONNECTIN G THE PCE-IR 52
1. Connect the circular connector on the USB cable provided to the PCE-IR 52 sensor
2. Connect the USB A connector on the USB cable provided to an available USB 2.0
port
3. Double click on the Software desktop icon
NOTE Do not disconnect the
program
C ONFIGURATION
Setting Temperature Units
Temperature units can be set on the main temperature panel by pressing °C or °F.
Setting Output Processing
The temperature range for the 4 to 20 mA output, the averaging period and the peak/
valley hold processing can be set by using the main program menu to select:
[Setup]
[Output Processing]
The temperatures corresponding to 4 and 20 mA can then be set (minimum –40°C,
maximum 1000°C, minimum span 100°C, maximum span 1000°C).
To minimise the the e ects of temperature
signal, the averaging period can be set between 0 and 60 seconds.
If required, Hold Processing can be applied by setting hold mode to Peak Hold or
Valley Hold and setting a hold period between 0 and 1200 seconds.
Setting Target Emissivity
To set an Emissivity value for the PCE-IR 52 Sensor, use the main program menu to
select:
[Setup]
[Emissivity]
Emissivity can then be set to a value between 0.1 and 1.0, or selected from a pre-
set list of materials by selecting [Data].
S ensor to the PC before the software is
S ENSOR TO A
PC
USB cable before closing down the
is:
Software
noise etc on the output
7

Publicidad

Tabla de contenido
loading

Tabla de contenido