Aesculap
®
Acculan 3Ti
Acculan 3Ti small drill GA671
Validated cleaning and disinfection procedure
Validated procedure
Manual cleaning and
wipe disinfection
16
Special features
■
When cleaning instruments with
movable hinges, ensure that these
are in an open position and, if
applicable, move the joint while
cleaning.
■
Drying phase: Use a lint-free cloth
or compressed air for medical
purposes
■
Ensure that the product is posi-
tioned in such a way that water
will not enter the product e.g.
through coupling interfaces.
(Immediately remove any fluid
that entered the product inadver-
tently.)
■
For cleaning the cannulation, use
special cleaning brush
(TA011944).
■
Clean difficult-to-access areas
with soft round plastic brushes of
fitting diameter.
Reference
Chapter Manual cleaning/disinfecting and
subsection:
■
Chapter Manual cleaning and wipe dis-
infecting