Only Philips Respironics approved parts may
be used to service this equipment. Use of non-
approved parts may invalidate the compressor's
test certificate and warranty.
We recommend you have your Porta-neb
serviced every 12 months by Philips Respironics
or a suitably qualified engineer.
8 Technical information
8.1 Product specification
Voltage
• 230 V~ 50 Hz
Power
• 210 VA
Fuse (in unit)
• T2 A / 250 V
Approximate dynamic pressure & flow*
• 7 L/min at 1 bar
Weight
• 2.9 kg
Dimensions
• L = 380 mm, W = 220 mm, D = 120 mm
Technical information
12196_RRDD_PortnebInst_1070641_REV2_v14.indd 11
Operating conditions:
Temperature
• Min. 10°C; max. 40°C
Air humidity
• Min. 30%; max. 75%
Atmospheric pressure
• Min. 700 hPa; max. 1060 hPa
Storage conditions:
Temperature
• Min. -25°C; max. 70°C
Air humidity
• Min. 10%; max. 95%
Atmospheric pressure
• Min. 700 hPa; max. 1060 hPa
* Pressure at the nebulizer jet and flow through
the nebulizer jet.
Fuse
You may need to change the fuse (T2 A / 250 V)
in the Porta-neb. This can be found in the
mains socket on the side of the Porta-neb.
Use a small screwdriver to release the panel
and then change the fuse (see diagram below).
W
D
Servicing manual/wiring diagrams
Full service manual/wiring diagrams are
L
available from Philips Respironics to help
qualified people service the Porta-neb
(part no. 1071642).